SETUP AND CALIBRATION, Cont.
End-Of-Print Line Feeds
At the end of a data transmission to a printer, the scale can send a number of line feed
commands to space the paper in the printer to the desired position for withdrawal or for the
next print. NOTE! This prompt will only be displayed if you select 3UW 1.
With the display showing (R3 , press the PRINT key to show the current setting. If
acceptable, press the PRINT key to save it. Otherwise use the UNITS key to select the new
setting and then press the PRINT key to save it. NOTE! This setting may be revised without
having to enter the calibration mode. Allowable values are: 00 through 99.
Setup And Calibration Is Completed
The setup and calibration process has been completed. The scale will reset and then display
weight. Replace the calibration screw removed earlier. The scale is ready to begin normal
operation.
SETUP REVIEW
The 6745 Baby Scale allows several operational parameters to be reviewed and changed as
necessary without having to enter the setup and calibration mode. The parameters in the
setup review will be processed in the following sequence:
1L&$G
0 = Disable battery charging (use Alkaline batteries)
1 = Enable battery charging (use Ni-Cad or NiMH)
2 = Enable battery charging and Force charging of discharged
Ni-Cad or NiMH batteries.
W+U6K 
Enable or Disable the threshold lock and select the weight that the
load on the scale must rise above to begin displaying weight
382
Enable or Disable automatic reset of weight display to zero when
scale is turned on
$6+
Disable or select number of minutes for automatic shutoff timer
6/3
Disable or select number of minutes of inactivity at zero for sleep
mode
%$8G
Select baud rate for serial printer port.
3UW< 
Select the serial data format.
3UW
Select the ticket format to be used when the PRINT key is pressed.
(R3
The Number of Ending Linefeeds Printed.
To enter the setup review mode:
1. Press the OFF key to turn the scale off
2. Press and hold the PRINT key and then press the ON key.
3. The display will then prompt with QL&$G, the selection to use Ni-Cad (NiMH) or Alkaline
batteries.
4. Refer to the instructions listed in the Setup and Calibration section of this manual for
information on how to change these parameters.