INSTALLATION, CONT.
Serial I/O Port
The 758C RS-232 serial I/O port is a 9-pin "D" connector on the rear
panel of the indicator. Figure No. 3 shows the Serial I/0 connector
along with the identity of the pins used. Make certain that the pins
are correctly identified before soldering a wire to them. Use t
connector retaining screws to hold the serial cable connector securely
to the rear panel.
5
9
4
8
3
7
2
6
1
Figure No. 3
he
PIN NO. FUNCTION
2 DATA INPUT (RXD)
3 DATA OUTPUT (TXD)
5 SIGNAL GROUND (GND)
Note that pins 1, 4 and 6 through 9 are not used.
The 758C RS-232 serial interface can be configured during the setup
and calibration procedure or during the setup review operation. Using
either method, it is possible to select the operation of the serial
interface as well as select the baud rate.
x The baud rates supported are: 1200, 2400, 4800, 9600, 19.2K
and 38.4K baud.
x The data format is fixed at 8 bits, No parity, and 1 stop bit.
x The indicator is shipped from the factory with the baud rate set
to 9600 baud.
Ethernet (Optional)
The 758C Ethernet option requires the purchase of the optional 758-IP Ethernet Interface
Option. Once installed, connections to the 758C are by CAT 5 or equivalent cable. Consult
your network administrator for the proper procedure to terminate and connect the cable.
BATTERIES
The 758C indicator can use 6 "C" size Alkaline, Ni-Cad or NiMH batteries (not included). You
must first obtain and install batteries before operations can begin. Batteries are contained in a
battery holder inside indicator. Access is via a removable panel on back of indicator.
CAUTION! The 758C can be operated from Alkaline, Ni-Cad or NiMH batteries. All
six (6) batteries must be of the same type. They must all be Alkaline, all Ni-Cad or
all NiMH. DO NOT mix Alkaline and Ni-Cad or NiMH batterie
s.
IMPORTANT! The 758C does
not have a battery charging circuit. Should you wish
to use Ni-Cad or NiMH batteries, they must be fully
charged before installing. When
discharged, Ni-Cad or NiMH batteries must be removed and placed in an external
charger to recharge.
CAUTION: RISK OF EXPLOSION IF BATTERY IS REPLACED BY AN INCORRECT
TYPE. DISPOSE OF USED BATTERIES ACCORDING TO THE INSTRUCTIONS.
ATTENTION: RISQUE D'EXPLOSION SI LA BATTERIES EST REMPLACE'E PAR UN
TYPE INCORRECT. REJETEZ LES BATTERIES UTILISE'ES SELON LES
INSTRUCTIONS.